- Anti-ProBNP dengan Ambang Batas 125 pg/L untuk Gagal Jantung
- Troponin I Membagi Pasien Menjadi Rule-In dan Rule-Out
- Metode CLEIA dan Alpha Hampir Sempurna dengan Korelasi 0,96-0,98
- Laboratorium Masa Depan Harus Sederhana, Akurat, dan Berbiaya Rendah
Full Transcript
High sensitive troponin eye and antibnp is important for manage the patient with cardiovascular disease importantly in
patient with an stemy and a heart failure. The value for high sensitive
failure. The value for high sensitive troponin eye will divided the patient into rule in observation and roll out
and will influence the manage of the patient.
I'm drina Sidarta. I'm clinical
pathologist at teaching hospital Muardi General Hospital in Solo Central Java. I
also as a head of clinical pathology laboratory at that hospital and I'm measuring in clinical infectious disease.
High sensitive stropanin eye and antiprnp is important for manage the patient with cardiovascular disease. According to the
cardiovascular disease. According to the ASC guideline to 2023 they recommended high sensitive troponin eye for rule in
and rule out patient with stemy.
cut depend on each hospital and method used in that hospital and that guideline also
recommended antib for their heart failure with the cut off of 125
picogram per liter importantly in patient with an stemy and a heart failure the value for High sensitive
troponin eye will divided the patient into rule in, observation and rule out and will influence the manage of the patient.
In my study, I compare the high sensitive troponin eye and antiprobnvine ray with the clea and the alpha method.
The correlation between the two method Clea and Alpha is very good with the correlation 0.96
and 0.98 almost 1.0.
So the correlation is very very good between the two method and the CV for the estroponin I and antiprob is very
good. The CV for the two method is below
good. The CV for the two method is below the CV marks allow and the bias from this two method also is very good for
the two method to I summarize my expectation for future laboratory is simple accurate and low
cost simple so everyone can handling technology accurate technology have to
give accurate result can differentiate between healthy and sick patient low cost because many government policy
nowadays they budget efficiency.
accurate have we can do in my laboratory Baik